Plate images from a run
A run can render the microscope image of a field of its CR-39 plate: the pits its alphas etch, seen through the microscope settings of the run. AlphaTrack counts such an image like any other, and the run then scores the count against what it knows the image holds.
Pits, optics and defects
The image is computed from the pits of the chosen field. Their size and shape follow from the energy and the angle of each alpha and from the etch. The microscope settings of the run set the pixel size, the numerical aperture and the wavelength, the focus, the illumination and the camera noise. Plate defects and background tracks are added at illustrative densities unless the run asks for a clean field.
How the image is labelled
Every rendered image says that it is an illustrative simulation without detector calibration, in four places: its file name starts with SYNTH_, the text fields of the PNG file carry the label, its metadata file states it, and the label stands beside the image on screen. The label goes with the image into AlphaTrack, into the run package and into a YieldProof bundle.
The answer table
The run keeps a table of every pit and defect in the image: where it lies, its size and the alpha that made it. The table is kept apart from the image and never reaches AlphaTrack's analysis; it is read only afterwards, to score the count.
Limits of the image
Until a CR-39 batch is calibrated with exposures to alphas of known energy, the response of the plate is illustrative, and so is the image: it shows what the model expects from the inputs, not what a given plate looks like under a given microscope.
The same inputs and seed give the same image. Changing only a microscope setting renders the image again and keeps the pits.